Beamline 8.3.1
Macromolecular Crystallography (TomAlberTron)
Current status | Operational; open to general users |
---|---|
ALS-U Status | This beamline may temporarily close ~2 months before the ALS-U dark time and will resume operations after a commissioning period. For more information, see ALS-U Beamline Impacts and other ALS-U web pages. |
Source | Superbend |
Minimum energy (eV) | 5,000 |
Maximum energy (eV) | 17,000 |
Wavelength range (Å) | 0.73–2.5 |
Research areas |
|
Technique category |
|
Flux/Brightness | 1.0 x 1012 photons/s measured at 3 x 0.35 milliradian divergence at 11,111 eV with ALS at 500 mA |
Resolving power | Water-cooled flat double Si(111) Khozu monochromator: E/ΔE = 7,000 measured with Ge(11,1,1water-cooled silicon bent into parabola 200 nanoradian slope ) analyzer crystal Primary Mirror: flat internallyerror Secondary Mirror: uncooled cyllindrical silicon bent into torroid 2:1 demagnification Both mirrors coated with Pt/Rh |
Access modes |
|
Additional notes | Primary Detector: Dectris Pilatus3 S 6M Secondary Detector: Hitachi Vortex EX energy-resolving for MAD and excitation scans |
Primary contact(s) | James Holton George Meigs |
Beamline website | http://bl831.als.lbl.gov/ |
Beamline phone number | 510-495-2083 |
Endstation 8.3.1
Tom-Alber-Tron
Status | Operational; open to general users |
---|---|
Research areas |
|
Technique category |
|
Spot size | 0.08 x 0.06 mm (v x h) |
Sample information | Sample temperature adjustable from 100-300K "Cool Hand Luke" Robot accepts all major pin types (Hampton, Yale, SPINE, etc.) and lengths (10-24 mm) and supports "delayed data collection mode" where data collection runs are deferred to the end of the shift. This robot does not work directly with cassettes because hand-mounting of samples in the mini-hutch is faster than any robot. |
Additional Notes | User-changeable motorized collimators: 100, 75 50, 30 20 microns diameter Goniometer is fast air-bearing spindle capable of "true" MAD with round-robin inverse beam and wavelength changes Sample Manipulator Huber XYZ stage 1.1 micron rms sphere of confusion |
Endstation 8.3.1 In-Situ
In-Situ Tray Goniometer
Status | Operational; limited access |
---|---|
Research areas |
|
Technique category |
|
Sample information | All SBS compliant crystallization trays supported Thin-walled in-situ ready trays such as in-situ-1 highly recommended Ambient temperature only currently supported |